9392-10B
Replacement pH Electrode 9392-10B
Replacement pH Electrode model 9392-10B is a replacement pH electrode for our lock-and-load systems.
Replacement pH Electrode Model 9392-10B Introduction
Our replacement pH electrode model 9392-10B is designed with a longer pH electrode length, to be used in our lock-and-load systems.
The electrode itself is a double junction pH Electrode with a peripheric annular non-fouling reference junction. The glass bulb is manufactured from our standard glass and offers excellent response time and precision measurement.
Replacement pH Electrode Model 9392-10B Features & Benefits
The 9392-10B replacement pH electrode is specifically designed to be a longer electrode for use with our lock-and-load retractable pH holder. These are used in conjunction with our pH controller range for industrial applications. While this pH electrode can be used in laboratory applications, we would suggest using our double junction pH electrode model 9092-10B, which has a shorter length for easier handling.
The 9392-10B is a sealed combination double junction glass replacement pH electrode, with our standard low impedance glass bulb, perhipheric annular ceramic reference junction and a silver - silver chloride element.
The 9392-10B offers precision readings; excellent repeatability; a rapid response time; is resistant to electrode poisoning with it's double junction and the annular peripheric ceramic reference junction is non-fouling. The electrode is also resistant to electrode leaching.
Where the 9392-10B is acting as a replacement pH electrode, for use in the stainless steel lock-and-load retractable pH electrode holder, we offer a stainless steel sheath. Please use model number 9392-10B-SS for the replacement pH Electrode with Stainless Steel Sheath.
The mounting cap on our replacement pH Electrode model 9392-10B is Ryton with a small flange for mounting the pH electrode into our PES style holders.
Glass pH Electrode Model 9392-10B Connection Types and cable length
The 9392-10B replacement pH electrode is available with a number of connection types. As standard, we supply the 9392-10B with 3 metres of connecting cable and a BNC fitting.
For custom fitting types please see the below table for the ordering number and options:-
-
Specifications
Replacement pH Electrode
Model No:- 9392-10B
Range:- 0-13 pH
Operating Temperature:- 0-100ºC
Electrode type:- Replacement pH electrode for the lock-n-load inline pH assembly
Junction:- Double Junction
Reference Type:- Non-fouling Peripheric Annular Ceramic with Silver - Silver Chloride reference
Dimensions:- 210mm long x 12mm ø
Additional:- A long bodied replacement pH electrode for the Lock-N-Load inline pH electrode system with reference gel, double junction and non-fouling reference.
-
Data Sheets
-
Ordering Information
Model Number Cable Length Connector 9392-3B 1 Metre BNC Connector (Most Common) 9392-10B 3 Metres BNC Connector (Most Common) 9392-S8 N/A S8 Connector (No Cable) 9392-3T 1 Metre Stripped & Tinned Connections 9392-10T 3 Metres Stripped & Tinned Connections 9392-3A 1 Metre U.S. Standard Connector fitted to older U.S. pH Instruments 9392-10A 3 Metres U.S. Standard Connector fitted to older U.S. pH Instruments 9392-3I 1 Metre Din 19262 Connector for German pH Instruments 9392-10I 3 Metres Din 19262 Connector for German pH Instruments 9392-3K 1 Metre Belling Lee Connector for older U.K. pH Instruments 9392-10K 3 Metres Belling Lee Connector for older U.K. pH Instruments For more information on the connection type please take a look at our pH electrode connection types page.
Electrodes with a longer cable or custom cable length are available to order.