call now 01785 254 597

Conductivity Dip cell range for immersion into tanks and other vessels

 

Dip Conductivity Cell

A dip Conductivity cell a conductivity cell designed to be immersed into an open tank or other unpressurised vessel.

Our dip conductivity cells are available with a wide range of conductivity cell constants varying from K=0.1 to K=10.0 for use with clean water and process solutions respectively. For our conductivtiy dip cells we recommend using a K=0.1 cell for a conductivity range of 0 - 10mS, K=1.0 for a range of 0 - 100mS and a K=10.0 cell for use with a range of 0 - 1000mS depending upon the conductivity instrument in use.

High Linearity Dip Conductivity Cells

Our D1 conductivity cells have specially engineered cell surfaces to optimise for linearity. Combined with the polarisation frequency from our conductivity controllers this ensures that the conductivity readings are highly linear across the measuring range of 5.00µS through to 1000mS depending upon the cell constant and instrument.

Our D2 conductivity cells are standard cells and available with a choice of riser tube, constructed in either PVC, cPVC, Stainless Steel or PVDF upon request. Depending upon the chosen cell constant these offer a conductivity range up to 50mS.

Dip Conductivity Cells With Automatic Temperature Compensation

All of our Dip Conductivity cells are available with automatic temperature compensation to ensure accurate conductivity measurement where  fluctuations in the temperature of the measured solution can occur. When ordering dip Conductivity cells with temperature compensation please specify the instrument that you're planning to use so we can ensure compatibility.

Dip Conductivity Cell range


D1-10

D1-10

The D1-10 Dip Conductivity Cell is a high linearity conductivity cell with a cell constant of K=1.0.

As a high linearity conductivity cell - the D1-10 will follow very closely a changing conductivity value giving very precise and readings across a full range of values.

The D1-10 is designed for a conductivity range of ~ 0 - 100mS (100,000)µS and is used in most general applications.

D1-10 is available in cPVC or stainless steel, with optional temperature compensation and mounting flange.

View Product


D1-100

D1-100

The D1-100 is a high linearity conductivity cell with a K=10.0 cell constant.

The D1-100 is designed for very accurately measuring fluids with high conductivities and has an operational range of 0 - 1000mS (1'000'000µS) and is often used in rinse water control, solution strength control and other process applications.

The D1-100 is available in cPVC and Stainless Steel, and has temperature compensation and mounting flange as options.

View Product


D2-01

D2-01

The D1-01 Conductivity Cell is a dip conductivity cell with a cell constant of K=0.1 for greater accuracy at the lower conductivity ranges.

Manufactured in PVC the D1-01 is designed for use in tanks and vessels which use clean water processes and where the conductivity doesn't exceed 2000µS (2mS).

Typically manufactured in PVC the D1-01 is available with optional temperature compensation and mounting flange.

View Product


D2-10

D2-10

The D1-10 Conductivity Cell is a dip conductivity cell with a cell constant of K=1.0 for use in most general applications.

The D1-10 is designed for use in tanks and vessels containing process solutions, such as pickling tanks, process rinse water, and detergent strength control, and where teh conductivity doesn't exceed 100mS (100'000µS).

Typically manufactured in PVC the D1-10 is available with optional temperature compensation and mounting flange.

View Product